| General Purpose RTDs |
|
|
|
|
Models 101, 102, & 103 Designed for a wide range of applications where a simple mounting scheme is required.
![]()
|
| 1. Model | Base Model/Series Number. |
| 2. A. Accuracy:
B. TCR: | Standard Class B (no code) High Class A (code H) Special Customer Specified (code S) * Industry Standard is DIN Curve (code 01B), Platinum, 100 ohms @ 0°C. Conforms to IEC 751, DIN 60751. Temperature Coefficient of Resistance is the temperature vs. resistance characteristics of a given metal (Pt, Cu & Ni) used in manufacturing the RTD. Determines the curve of the RTD. Ro - Resistance at 0°C (32°F) Dependent on sheath diameter, the smaller the diameter - the faster the response. See RTD General Specs. Model 102 Copper Tip, element is encapsulated in copper to increase sensitivity at tip of probe. |
| 3. Construction: | Code A - 316SS tube and wire construction, thin film element (.00385055) TCR, Teflon insulated lead wire. Code C - 316SS tube and wire construction, wire wound element, fiberglass insulated lead wire. Code B & D - Inconel sheathed MgO construction, wire wound element, fiberglass insulated lead wire. |
| 4. Lead Wires: | ![]() |
| 5. Sheath Diameter: | .250” (1/4”) is the industry standard. Code C/N for reduced tip design. |
| 6. Sheath Length: | Entire stem length. |
| 7. Lead Wire Length: | Length of wires beyond the sheath. Use Z000 if connector is installed directly on the end of sheath. |
| 8. Lead Wire Protection: | Stainless Steel Overbraid or Stainless Steel Armor. Order length at 3”-6” shorter than lead wire length. |
| 9. Connectors: | Optional male plug for sensor end. For mating jack add code “/J”. |
| 10.Optional Fittings: | Sensor is not supplied with mounting hardware, specify optional fitting. |
| 11.Water resistant: | Increases moisture protection for humid environments. |
![]() |